Waveform Digitizers 7 Analog Modules
System Reference, January 2001
213
lution digitizer for another test item for the Aout pin.
When K1 and K3 are closed and K2 is open, the signal from
the Aout pin can be measured by the high speed digitizer.
When K1, K2, K4, and K5 are closed and K3 is open, the
signal from the Aout pin can be measured by the high reso-
lution digitizer. Use of the loop back route can make the
measurement performance better than when user relays
are used on the DUT board for changing the high speed
digitizer and high resolution digitizer.
Figure 96 Example for Using Loop Back Route
A+
A-
B+
B-
C+
C-
D+
D-
SYNC CLK
loopex
SYNC DATA
Conversion
Clock
Start/
End
ADC
Filter
Sequencer/
Waveform
Memory
Timing
Generator
Input
Amp
+
-
DC
offset
A+
A-
B+
B-
C+
C-
D+
D-
SYNC CLK
High Speed Digitizer
SYNC DATA
Conversion
Clock
Start/
End
ADC
Filter
Sequencer/
Waveform
Memory
Timing
Generator
Input
Amp
+
-
DC
offset
Aout
DUT
High Resolution Digitizer
K1
K2
K3
K4
K5
When K1, K3 are closed and
K2 is open.
When K1, K2, K4, and K5 are
closed and K3 is open.