Synchronization 7 Analog Modules
System Reference, January 2001
255
The following describes the details of how to remove the
synchronization uncertainty when using same master
clock for a trigger source and an analog module.
You can do either of the following to remove uncertainty
that occurs in the trigger-to-signal delay of the analog
module.
NOTE You can add the specified delay time value to the trigger-
to-signal delay for each analog module. It is useful to
adjust the edge location of the trigger signal to the desired
position.
Program the edge location of the trigger signal at the
beginning of the tester period and inform the system of
the trigger line length on the DUT board by using the
TRGL firmware command.
Every beginning of the tester period fully synchronizes
with the master clock. At the analog module, the phase
information between the master clock edge and the
trigger edge is obtained by treating the beginning of the
tester period as the reference point. The TRGL
command calculates the difference of the trigger line
length at the timing calibration and your trigger line
length on the DUT board, then compensates the phase
between the master clock edge and trigger edge, by
treating the beginning of the tester period as the refer-
ence point. Therefore, you have to program the edge
location of the trigger signal at the beginning of the
tester period. The following is the timing chart.
NOTE You need to specify the trigger line length with ±200 ps
(typical) accuracy in the TRGL firmware command to
remove the uncertainty.