AWG G3 PAD FOR PS3 Dokumentacja Strona 241

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 285
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 240
Time Interval Analyzer 7 Analog Modules
System Reference, January 2001
241
Figure111 TIA Instrument Block Diagram
The TIA instrument has two signal inputs and one trigger
input. Each signal input can be routed to the internal
paths for the start and stop signals. This enables events
such as pulse width or pulse period to be measured on a
single input with one channel only.
Two signal inputs must be used for measuring propaga-
tion delay.
The trigger signal determines the measurement start
timing as an arming trigger.
Channel 1
Channel 2
Trigger
Time Measurement
Block
TIA
Start path
Stop path
Przeglądanie stron 240